Classification of air quality for manufacture of sterile products

Clean room classes

A

B

C

D

Comparable clean room class according to...

     

US Fed. Std.209

100

100

10,000

100,000

VDI 2083

3

3

5

6

Type of air flow

 

Laminar flow in the working area

Non-laminar flow

Non-laminar flow

Non-laminar flow

Air change

1/h

360540

20

20

20

Average air velocity

m/s

       
 

vertically

0.3

 

horizontally

0.45

Maximum permitted number of viable microorganisms

1/m3

1*

max. 5*

max. 100

max. 500

Maximum permitted number of particles

1/m3

       
 

0.5 mm

3,500

3,500

350,000

3,500,000

 

5 mm

none

none

2,000

20,000

The marked low values can be reliably determined only by a great number of air samples.